| Model Number |
3300 |
| Medical Device Problem Codes |
Defective Device (2588); Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory; functional testing and baseline checks were completed.Visual inspection revealed stains on the front and back laramie.The programmer was powered on and the logfile was downloaded; data review revealed that no error/fault codes had been recorded.No error or fault codes were recorded in the programmers logfile and benchtop testing was unable to reproduce the behavior.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.
|
| |
|
Event or Problem Description
|
|
It was reported that the touchscreen of this programmer was unresponsive.This programmer was returned for analysis.
|
| |
|