| Model Number |
WD 290 IQ |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A metal clamp intended to fasten the water drain tube to the device was tightened excessively during assembly, cutting the tube material.The cut was not visible, being covered by the clamp itself.After seven months of routine use, the tube started to leak.The damaged component has been replaced and the device is functioning normally.
|
| |
|
Event or Problem Description
|
|
The plastic drain of the washer was damaged by the hose clamp, producing a deep cut that started to leak and flooded the floor of the cssd around the device.
|
| |
|