| Model Number |
3300 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the touchscreen of this programmer was damaged.The programmer was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Visual inspection of the device revealed that the device does not detect the touch screen and there was gap on assy foot pad.The programmer was powered on and it was confirmed that the touchscreen was intermittently unresponsive due to not detecting touch inputs.No error or fault codes were recorded in the programmers logfile.The programmer was disassembled, and it was determined that an issue with the screen's touch controller caused the observed touchscreen behavior.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.
|
| |
|