Model Number X060-1520
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was made aware of a product complaint on 5/04/2023.It was reported that a system inserter/compressor was broken when the incorrect set screw of the instrument was rotated for final tightening of a system implant.The complainant reported that a system screwdriver was also broken during the inserter/compressor malfunction.There were no known patient complications or delay in treatment associated with this complaint.An alternate available inserter/compressor and screwdriver were used to successfully complete the procedure.An rma# was issued for return of the complaint instruments, which were received at the manufacturer for assessment on 5/17/2023.This report is associated with mdr# 3005031160-2023-00009.
 
Additional Manufacturer Narrative
A visual assessment of the returned system inserter/compressor showed an instrument with minimal use, as identified by crisp laser markings and a lack of surface scratches.The distal lip of one arm of the inserter/compressor was broken and not present.A functionality assessment was not performed due to the damaged condition of the returned system inserter/compressor, which was removed from distributable inventory.A dhr review was performed for the complaint inserter/compressor lot# and there were no manufacturing anomalies identified.The instrument lot met all required specifications prior to being released to distributable inventory.The complaint instrument lot has been available for distribution since 6/14/2019.It was reported that the system inserter/compressor was damaged when the incorrect set screw was tightened by the surgeon.There are two grasping arms of the inserter/compressor that engage with the system cross bar plate and system locking plate.The correct set screw to tighten after final placement is laser marked on the correct arm of the inserter/compressor with " final locking", which tightens the system locking plate onto the system cross bar plate.The force required to damage the distal lip of the inserter/compressor also contributed to the distal tip of the system screwdriver malfunction.There have been three other complaints of similar nature in the past 12 months.The manufacturer will continue to monitor these instruments for complaints from the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXLE INTERSPINOUS FUSION SYSTEM
Common Device Name
SPINOUS PROCESS PLATE
Manufacturer (Section D)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT
Manufacturer (Section G)
XTANT MEDICAL HOLDINGS, INC.
664 cruiser lane
belgrade MT
Manufacturer Contact
rebecca lennemann
664 cruiser lane
belgrade, MT 
MDR Report Key17050195
Report Number3005031160-2023-00008
Device Sequence Number3377696
Product Code PEK
UDI-Device IdentifierM697X06015201
UDI-PublicM697X06015201
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K130438
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX060-1520
Device Lot Number14652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2023
Initial Report FDA Received Date06/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age20 YR
Patient SexUnknown
-
-