• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 REVERSE LINER TRIAL 32+0 R; SHOULDER INSTRUMENT - HUMERAL TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 REVERSE LINER TRIAL 32+0 R; SHOULDER INSTRUMENT - HUMERAL TRIAL Back to Search Results
Model Number 6500-32-100
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the case the scrub tech noticed that the 32 +0 r poly trial was missing a locking tab.Since we were unable to use the trial the doctor trialed with a different size to make his final implant decision.There was no negative impact to the patient and surgery time was not extended due to this.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE LINER TRIAL 32+0 R
Common Device Name
SHOULDER INSTRUMENT - HUMERAL TRIAL
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17051366
Report Number1818910-2023-11457
Device Sequence Number7039666
Product Code HWT
UDI-Device Identifier10603295547297
UDI-Public10603295547297
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6500-32-100
Device Catalogue Number650032100
Device Lot Number332253
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2023
Supplement Date Received by Manufacturer08/31/2023
Initial Report FDA Received Date06/02/2023
Supplement Report FDA Received Date08/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/16/2011
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-