| Model Number |
G408324 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This device was reported to have a bleed-back valve that would not hold pressure.Blood was reported to be flowing out of the end of the agilis device until another catheter was re-inserted.No injuries were reported.The date of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Innovative health, llc became aware on (b)(6) 2023 of an agilis nxt steerable introducer from (b)(6) hospital reported to have a back-bleed valve that would not hold pressure.Blood was flowing out of the end of the agilis until another catheter was re-inserted.The date of the event is unknown.Innovative health received the device for evaluation on (b)(6) 2023.Upon investigation, the device hemostasis cap was identified to be insufficiently adhered to the sideport hub.No injuries were reported.
|
| |
|