| Model Number |
3.0MM MGS DRILL BIT, STERILE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 05/09/2023, it was reported by a sales representative via sems that an ar-9628s drill bit broke off.This was discovered during a case, device breaks during use.Per facility: "the surgeon was drilling for a non-locking peripheral screw and believes he hit the posterior retractor, in which the tip ended up breaking off.After the baseplate was removed, he was able to retrieve the broken piece.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to hitting the device against another device.
|
| |
|