| Model Number |
657432 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "a straight plate of the variax foot system was required for a reduction of the distal tibia and at the time of placing the screws to fix it, they went through the holes of the plate displacing the plate and it had to be removed and it was needed to find another way to maintain the reduction increasing the surgical time.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: refer to b2 outcomes attributed to ae.The reported event could not be confirmed, since the device was returned and found to be fully functional.Device inspection revealed the following: the received bone screw was visually inspected in which we can see the damaged head the shape of the screw head is clearly damaged, which is caused by misalignment and application of excessive force which can be implantation or explantation.A functional test was conducted with the received plate and screws in which we were able to assemble them properly.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation root cause can be attributed to user-related.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "a straight plate of the variax foot system was required for a reduction of the distal tibia and at the time of placing the screws to fix it, they went through the holes of the plate displacing the plate and it had to be removed and it was needed to find another way to maintain the reduction increasing the surgical time.".
|
| |
|