| Model Number |
ECS33A |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/5/2023.D4: batch # 452a98.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a rectum procedure surgery, the device could not fire.Changed to the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/30/2023 d4: batch #452a98 investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs33a device arrived with no apparent damage.The breakaway washer was present and uncut, the device was fully loaded with staples, indicating that the device had not been fired.The device was tested for functionality with the returned washer and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.However, one missing staple was found in the test media after firing the device.The remaining staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected and released to approved specifications.It should be noted that an inadvertent movement of the firing trigger after disengaging the safety might cause the staples to fall out.It is important to ensure that the adjusting knob is set in the firing range before disengaging the safety.Therefore, it is imperative that disengaging the safety be the last step prior to firing.It should be noted that all devices are inspected 100% for staple presence by an automated vision system and are visually inspected 100% as a final check.In addition, at finished goods the devices are visually inspected based on a sample.Please reference the instruction for use for more information.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: to remove the spacer tab, open the instrument by turning the adjusting knob counterclockwise two revolutions.Place purse-string sutures in the organs to be anastomosed.Based on surgeon experience and judgment, a closed lumen technique (double or triple stapling technique) may be employed as an alternative to a purse-string technique.For a double stapling technique, open the instrument using the adjusting knob until the orange tying area is visible.Remove the anvil to expose the trocar.Retract the trocar by rotating the adjusting knob clockwise until a stop is reached.Check trocar to verify that it is fully retracted before proceeding.Insert the anvil into the lumen and secure the purse-string onto the anvil shaft above the tying notch.Insert the device up to the closed lumen with the anvil removed and the trocar fully retracted.Fully extend the trocar and pierce tissue by gently rotating the instrument while holding the adjusting knob.Push the tissue down until the orange tying area is visible.While closing the instrument, keep the organ segments in proper orientation.Inspect to ensure extraneous tissue is excluded.Tissue should lay flat and be evenly distributed within the instrument.Turn the adjusting knob clockwise to close the instrument.As the final adjusting revolution is approached, the orange indicator moves into the green range of the tissue compression scale.The event described could not be confirmed as the device performed without any difficulties.A manufacturing record evaluation was performed for the finished device batch number 452a98, and no non-conformances related to the reported complaint condition were identified.
|
| |
|
Search Alerts/Recalls
|
|
|