| Model Number |
201-D |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the field, this was a coil embolization procedure for unruptured basilar tip aneurysm.The procedure was performed by double catheter technique.A pulserider t, 3mm, 8mm arch (201d, 3078569818) was used with the prowler select plus 150/5cm microcatheter (606s255x, 30883124), while the phenom17 was used for the coil.After the prowler select plus was inserted to the aneurysm, the pulserider was attempted to inserted, but there was a resistance, so the pulserider was withdrawn.The issue was resolved by replacing both the pulserider and the prowler select plus to new ones.The coil embolization procedure was completed with successful pulserider implantation.There was no negative impact to the patient.A continuous flush was not done.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section e1 - initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 3078569818 number, and no non-conformances related to the malfunction were identified.The company is seeking information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2023-00357.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Section b5: additional information received.Indicated, that the device was difficult to advance.The prowler select did not kink or bend.There were no damaged to the pulserider.Section e1.: initial reporter phone: (b)(6).Complaint conclusion: as reported, by the field.This was a coil embolization procedure, for unruptured basilar tip aneurysm.The procedure was performed, by double catheter technique.A pulserider t, 3mm, 8mm arch, (201d, 3078569818) was used with the prowler select plus 150/5cm microcatheter ((b)(4)), while the phenom17 was used for the coil.After the prowler select plus was inserted to the aneurysm, the pulserider was attempted to inserted, but there was a resistance, so the pulserider was withdrawn.The issue was resolved by replacing both the pulserider and the prowler select plus to new ones.The coil embolization procedure was completed with successful pulserider implantation.There was no negative impact to the patient.A continuous flush was not done.Additional information received.Indicated, that the device was difficult to advance.The prowler select did not kink or bend.There were no damaged to the pulserider.The device was discarded.Therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 3078569818 number.And no non-conformances related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication, that the event is related to the device manufacturing process.The exact cause of the event could not be determined.However, there are circumstances of the procedure that may have contributed to the reported failure.Per the instructions for use (ifu), never advance or apply torque against resistance without careful fluoroscopic assessment of the cause.Movement against resistance may result in damage to the vessel or the device.If the cause cannot be determined, withdraw the device from the patient and replace with a new device.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since, there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|
|