| Catalog Number |
82009 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that 30k cav.Thinsert-fg,packed got hot and had no water coming out of tip during use.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Adding udi # - (b)(4).This is a follow up report to add the udi#.Returned qty.1 insert, 82009, 22021-00068765, 40, beyond useful life test method / gp005-wi02 inductance: gauge id# 5349 due: 11/30/2024 result: 2.99 ¿ meets spec ¿ does not meet spec ¿ n/a tip condition: - ¿ meets spec ¿ does not meet spec ¿ n/a stack condition ¿ meets spec ¿ does not meet spec ¿ n/a tested on: g136 gage id# 9626-2 due date: 09/30/2023 set pressure: 35 psi water flow: output rate: 35 psi - ¿meets spec ¿does not meet spec ¿n/a spray pattern: - ¿ meets spec ¿does not meet spec ¿ n/a water leak: ¿ hp sealing ring ¿proximal ring ¿distal ring ¿seam leak ¿n/a vibration: - ¿meets spec ¿does not meet spec ¿ n/a grip condition: ¿ meets spec ¿ does not meet spec ¿ n/a other visual observation: insert laminate stack is bent.Also insert was tested on a digital thermometer i.D.#6116a due date: 09/30/2023.The temperature was 82.9° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: ¿ confirmed - ¿ not confirmed.
|
| |
|
Search Alerts/Recalls
|