Catalog Number 82008
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
In this event it is reported that a 30k fsi-sli-fg-1000 ins, pkd heats up during use.No injury.
 
Additional Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Additional Manufacturer Narrative
Additional information added: udi # (b)(4).This is a follow up to report this information.Returned qty.1 insert, 82008, 23073-00093807, 30, warranty.Test method / gp005-wi02.Inductance: gauge id# (b)(4) due: 11/30/2024 result: meets spec does not meet spec n/a.Tip condition: - meets spec does not meet spec n/a.Stack condition meets spec does not meet spec n/a.Tested on: g136 gage id# (b)(4) due date: (b)(6) 2023 set pressure: 35 psi water flow: output rate: 35 psi - meets spec does not meet spec n/a.Spray pattern: - meets spec does not meet spec n/a.Water leak: hp sealing ring proximal ring distal ring seam leak n/a.Vibration: - meets spec does not meet spec n/a.Grip condition: meets spec does not meet spec n/a.Other visual observation: insert laminate stack has a big solder.Also insert was tested on a digital thermometer i.D.#(b)(4) due date: (b)(6) 2023.The temperature was 82.9° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FSI-SLI-FG-1000 INS,PKD
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key17059090
Report Number2424472-2023-00381
Device Sequence Number15581327
Product Code ELC
UDI-Device IdentifierD003820081
UDI-PublicD003820081
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K052334/K970
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number82008
Device Lot Number23073
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/02/2023
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/02/2023
Supplement Date Received by Manufacturer06/02/2023
Initial Report FDA Received Date06/05/2023
Supplement Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A