C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Codes |
Poor Quality Image (1408); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by a tm - the image is blurry, and the tip can't be found.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been received by the manufacturer for evaluation.H3 other text : device not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded.The device was returned to the service facility for evaluation.During evaluation, the reported issue of the image is blurry, and the tip can't be found was unconfirmed; the unit is giving a clear and bright image.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported by a tm the image is blurry, and the tip can't be found.
|
| |
|
Search Alerts/Recalls
|
|
|