| Model Number |
70805-70 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, product has not been returned for further investigation.The useful life of freestyle lite meter is 5 years.As the manufacturing date of this product is 2017, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Event or Problem Description
|
|
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned meter was investigated with retained test strips.Meter powered on with button depression and test strip insertion.No new issues were observed.Blank screen was not observed.The complaint was not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Search Alerts/Recalls
|