| Model Number |
4300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This incident is related to a skin irritation problem.Konix gel k101952 is a component of the device.There were no dysfunction, deficiencies/deteriorations found with the device.No device malfunction has occurred.The review has determined that the risk assessment is still adequate.
|
| |
|
Event or Problem Description
|
|
Information provided states that patient was prescribed accelstim device (b)(6) 2022.Pt said she only used it 3-4 times.After last treatment red burns appeared about the size of the device on her foot, 2-3 burns on the back bottom of her foot where the device was and on the top of the foot.She stopped using the device.Patient said the burns have since healed.Pt stated she tried alternative mineral oil and glycerin with the same results.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the returned device was completed on 08 may 2023.Five therapy sessions were conducted, with the last session occurring on (b)(6) 2023, lasting 10 minutes.During the treatment, device temperature was continuously monitored and never exceeded 25°c.Following the treatment, a burn-in test was conducted, which revealed no errors.Additionally, a force balance assessment was performed to determine ultrasound intensity value, and no deviations were detected.
|
| |
|