• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BENDING DRIVER F/ 3.5MM VA LOCKING HOLES; BENDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BENDING DRIVER F/ 3.5MM VA LOCKING HOLES; BENDER Back to Search Results
Model Number 03.221.251
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2023 where a rfna was inserted, was malleting in with sleeve in jig.Sleeves were not able to be drilled through because it hit nail and they were stuck.Unable to taken out - mallet also was chipped in process.Nail was not billed.There was no surgical delay.There were no patient outcome or consequences.This report is for one (1)bending driver f/ 3.5mm va locking holes.This is report 5 of 8 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENDING DRIVER F/ 3.5MM VA LOCKING HOLES
Common Device Name
BENDER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key17060338
Report Number8030965-2023-07078
Device Sequence Number5130143
Product Code HXW
UDI-Device Identifier10886982277022
UDI-Public(01)10886982277022
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.221.251
Device Catalogue Number03.221.251
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/19/2023
Initial Report FDA Received Date06/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILL SLEEVE / 4.2MM; DRILL SLEEVE / 4.2MM; PROTECTION SLEEVE/ 11/8; PROTECTION SLEEVE/ 11/8; RFNA/ 10MM/ 440MM/ STANDARD BEND/ STER; SPIRAL COMBINATION HAMMER 500 GRAMS
Patient SexUnknown
-
-