| Model Number |
1012631-59 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524); Difficult or Delayed Activation (2577); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the heavily calcified distal right iliac artery.After crossing with an unspecified wire the 8.0x59mm, the omnilink elite stent delivery system failed to fully cross lesion due to resistance with anatomy; therefore, it was attempted to pull back the delivery system into sheath.Resistance was met and the stent began to come off the balloon.The omnilink elite was advanced back into the iliac artery and deployed noting that it was also, partially deployed in the proximal right iliac artery.A smaller unspecified balloon was advanced into the portion of the stent that was deployed and ballooned enough to use another omnilink stent to fully deploy the first stent, while also stenting the rest of the lesion.There was no adverse patient sequela an no clinically significant delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.The investigation determined the reported difficulties and subsequent treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|