| Catalog Number |
06481647190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable mg2 magnesium gen.2 results for 3 patient samples on a cobas integra 400 plus.The initial results were reported outside of the laboratory.The doctor questioned the results and the samples were repeated.For patient 1, the initial mg2 result was 4.3 mg/dl.The repeat result was 2.0 mg/dl.The sample was repeated again using a new reagent pack and the result was 1.9 mg/dl.For patient 2, the initial mg2 result was 4.2 mg/dl.The repeat result was 1.8 mg/dl.The sample was repeated again using a new reagent pack and the result was 1.6 mg/dl.For patient 3, the initial mg2 result was 4.5 mg/dl.The repeat result was 2.4 mg/dl.The sample was repeated again using a new reagent pack and the result was 2.1 mg/dl.The first repeat results were deemed correct.The analyzer serial number is (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration was last performed on 26-apr-2023.The qc recovery data provided was acceptable.The alarm trace did not contain a conspicuous event.The field service engineer (fse) monitored the analyzer and confirmed it was operating as intended.No further issues were reported after the service visit.The investigation did not identify a product problem.The root cause of the event could not be determined.H3 other text : na.
|
| |
|