NUVASIVE INC NUVASIVE RELINE SYSTEM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS Back to Search Results
Model Number 10001139
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
During a spinal procedure at l3/pelvis, the tip of the driver was reported as fractured off clean and left in-situ in the left iliac screw.No adverse patient consequences reported.
 
Additional Manufacturer Narrative
The device was returned to nuvasive and the complaint was confirmed.Examination of the returned device was completed by nuvasive and the fracture pattern identified was consistent with excessive force fracture.Review of the information received suggests excessive off-angled force and or sclerotic bone as possible cause or contributors.The root cause was determined to be component failure related to excessive force and possibly the result of off-angled approach and or sclerotic bone.The unretrieved tip in the patient is not a concern for migration as it was reportedly firmly secured in the screw.No additional investigation required.Labeling review: ".Warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".Do not implant the instruments: complications to the patient may include, but are not limited to: breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration.Breakage could cause injury to the patient." ".Pre-operative warnings: the method of use for the instruments are to be determined by the user¿s experience and training in surgical procedures.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.Care should be used during surgical procedures to prevent damage to the devices and injury to the patient." ".Intra-operative warnings: the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient.Over-bending, notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage.The physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS
Manufacturer (Section D)
NUVASIVE INC
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE INC
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key17066230
Report Number2031966-2023-00124
Device Sequence Number6775493
Product Code LXH
UDI-Device Identifier00887517955432
UDI-Public887517955432
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10001139
Device Lot NumberEM46631
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2023
Initial Report FDA Received Date06/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-