INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; PROGRASP FORCEPS Back to Search Results
Model Number 420093-14
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the jaw of the prograsp forceps was not fully opened and was not moving smoothly.The customer used a backup instrument to continue.The procedure was completed with no reported injury.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the prograsp forceps instrument involved with this complaint and completed the device evaluation.Failure analysis investigations replicated and confirmed the customer reported complaint.This instruments grip tips were not moving intuitively, i.E.They were not following the master tool manipulator (mtm) input commands smoothly.The instrument's motion on the system was confirmed to be non-intuitive as the yaw and pitch movements were off by about 90 degrees.When the instrument was off the system, it was noticed that its neutral position was rolled about 45 degrees to one side rather than resting in the middle as it should.The housing was removed and the input gear tabs were found to have some light material scraped off.The tab was not broken or cracked.Manually rolling the instrument too far can cause the tabs to slip down allowing the instrument to roll without the input turning an equal amount.When the input pops back into place the instrument will longer have intuitive motion.To confirm this, the input gear tabs were purposely pushed down to disengage the gear.The input was then lined to its correct orientation and re-engaged.The instrument was again tested on a system and moved properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
PROGRASP FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key17067311
Report Number2955842-2023-15713
Device Sequence Number12294498
Product Code NAY
UDI-Device Identifier00886874111345
UDI-Public(01)00886874111345(10)N12191202
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420093-14
Device Catalogue Number420093
Device Lot NumberN12191202 962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2020
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/30/2020
Initial Report FDA Received Date06/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-