• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF CMNTLS IMPLANT INSERT; KNEE INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF CMNTLS IMPLANT INSERT; KNEE INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the partial left knee arthroplasty that the instrument fractured.No piece fell nor remain into the patient body.All pieces were found.No impact or injury to patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Associated products: item reference: (b)(4), item name: oxf uni ph3 tib impactor, item lot: zb7117655.Foreign- canada.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2023-00180.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.Visual inspection shows the item and lot numbers match the complaint.Product was not returned in its original packaging and there were missing parts.Device showed wear and damage and the tip was missing on the device.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.A review of the raw material certificates confirmed no abnormalities or deviations.Devices are used for treatment.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF CMNTLS IMPLANT INSERT
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key17069671
Report Number3002806535-2023-00181
Device Sequence Number6105090
Product Code LXH
UDI-Device Identifier00880304525672
UDI-Public(01)00880304525672(11)191104(10)ZB130404
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number32-422097
Device Lot NumberZB130404
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2023
Supplement Date Received by Manufacturer08/25/2023
Initial Report FDA Received Date06/06/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
-
-