| Model Number |
27040GP140-S |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was event with 27040gp140-s cutting loop, bipolar 24/26 fr.According to the information received from the salesrep: according to the facility the item broke off during the procedure and into the patients bladder.Patient harm was not noted at time of this report.Additional information has been requested but not yet received as of the date of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was received by manufacture, the complaint is confirmed the distal end is broken off and the yellow plastic is burnt at distal end temperature damage can be seen at proximal end.Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.The event is filed under internal karl storz complaint id (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per manufacture findings: the investigation revealed that the wire on the active and passive electrodes had melted in spots.This is an indication that there was a short circuit.Depending on the number of applications, the wire becomes thinner with each application, which could ultimately have led to the short circuit.On the other hand, a wire may also have bent, for example, so that the distance between the electrodes was too small, resulting in a short circuit.The event is filed under internal karl storz complaint id (b)(4).
|
| |
|