Model Number PHAC1254
Medical Device Problem Codes Break (1069); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, the patient underwent revision surgery, the broken femoral stem was identified, and soft tissue was removed around the femoral head and it was removed, we then removed the femoral implant.We implanted another stem.We took an intraoperative x-ray and noted there was a piece of the drill bit that must of broken during the osteotomy portion of the case and distal along the stem.We then placed the size a/50 mm body with the appropriate version and impacted it and placed a locking screw.We then placed a 36/-6 biolox head and impacted.Products not revised: product id: dsbfgg58, lot: 05113366701551930, qty: 1.Product id: 18080303, lot: 1543191, qty: 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® NECK VAR/VAL 8DG LONG COBALT CHROME
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
rachael wise
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key17072760
Report Number3010536692-2023-00111
Device Sequence Number6776039
Product Code JDI
UDI-Device IdentifierM684PHAC12541
UDI-PublicM684PHAC12541
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K091423
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPHAC1254
Device Catalogue NumberPHAC1254
Device Lot Number1577764
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2023
Supplement Date Received by Manufacturer05/17/2023
Initial Report FDA Received Date06/06/2023
Supplement Report FDA Received Date07/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
-
-