| Model Number |
5301000ND |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Electrolyte Imbalance (2196)
|
| Date of Event |
05/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device was received from (b)(6) memorial.The device error log was reviewed and it was noted that there were arrest occlusion alarms that showed to the user during the case.Based on the log data that was reviewed, it does not seem that the console caused or contributed to the complaint condition of high serum potassium.The console was found to have delievered the correct/expected amount that was intended to be delivered by the user based on the settings that were utilized during the case.Quest will continue to monitor this complaint condition.Quest has concluded its investigation of this device.
|
| |
|
Event or Problem Description
|
|
It was reported by (b)(6) memorial of an alleged issue with mps 3 nd unit.The unit was alleged to have repeatedly errored (ec 103) while trying to deliver cardioplegia.The unit was powered down and re-started with no success.After the final dose was given and the cross clamp was removed, a blood gas level was drawn and noted a high serum potassium level.It was confirmed that there were no patient complications.
|
| |
|
Search Alerts/Recalls
|