• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A CR BEADED #4L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A CR BEADED #4L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C Back to Search Results
Model Number 5517-F-401
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 05/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a triathlon femoral component was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the femoral component.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : device not returned.
 
Event or Problem Description
Dr.Revised a triathlon left cementless femur and cs insert due to femoral loosening.Felt bone quality was poor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON P/A CR BEADED #4L
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, C
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key17076287
Report Number0002249697-2023-00647
Device Sequence Number6329164
Product Code MBH
UDI-Device Identifier07613327040937
UDI-Public07613327040937
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2024
Device Model Number5517-F-401
Device Catalogue Number5517F401
Device Lot NumberHD29H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2023
Initial Report FDA Received Date06/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-