| Model Number |
682000 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a hemodynamic pressure monitoring procedure, the clinical staff noted that the pressure monitoring set was leaking fluid while connected to the patient.The device was replaced for the patient.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|