• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/10 MM R; TIBIAL KNEE INSERT FIXED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/10 MM R; TIBIAL KNEE INSERT FIXED Back to Search Results
Model Number 02.12.0210FR
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 05/11/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 15 may 2023.Lot: 1909427: (b)(4) items manufactured and released on 04-dec-2019.Expiration date: 2024-nov-23.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Additional item involved in the event, batch review performed on 15 may 2023.Gmk-sphere 02.12.0702r femoral component sphere tinbn coated size 2 r (k202684) lot: 174170 lot: 174170: (b)(4) items manufactured and released on 11-oct-2017.Expiration date: 2022-oct-01.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Preliminary investigation performed by r&d knee manager: revision surgery of a gmk sphere implant after almost 2 years from primary implantation due to instability.The liner was replaced with a thicker one but the surgeon will revise all the system to pass to an hinge in august this year, due to medial instability.During the revision surgery it was noted that the coating on the sensitin femoral component was 'rubbed off in one point'.Despite this damage, the patient and the other implanted components don't have any problems.From the picture of the implant sent for investigation, we can notice a stripe on the distal medial condyle where the coating is no more present.The defect seems to have a very defined and regular shape and looks different from scratches expected in case of detachment caused by a third body.No further considerations can be done just looking at the picture sent.
 
Event or Problem Description
About 1 year and 10 months after the primary surgery, the liner was revised with a thicker one (from 10 to 12mm) due to instability.During the revision, the tinbn-coating of the femoral component was found rubbed off in one condyle.No material was found between the femur and the insert.The femoral component was not revised.The surgeon plans to revise the patient to switch to a hinge system because the instability persists.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 2/10 MM R
Common Device Name
TIBIAL KNEE INSERT FIXED
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key17077576
Report Number3005180920-2023-00417
Device Sequence Number11306255
Product Code JWH
UDI-Device Identifier07630030826146
UDI-Public07630030826146
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K121416
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/23/2024
Device Model Number02.12.0210FR
Device Catalogue Number02.12.0210FR
Device Lot Number1909427
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2023
Initial Report FDA Received Date06/07/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/04/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-