| Model Number |
SD980.001 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Physical Asymmetry (4573)
|
| Date of Event |
11/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation ongoing; rc unknown.
|
| |
|
Event or Problem Description
|
|
Bsso plates removed due to patient's mandible as being asymmetric with suboptimal occlusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation concluded that the devices met specifications and did not malfunction.No clear root cause could be established for patient's asymmetric mandible with suboptimal occlusion.
|
| |
|
Event or Problem Description
|
|
Bsso plates removed due to patient's mandible as being asymmetric with suboptimal occlusion.
|
| |
|
Search Alerts/Recalls
|