| Catalog Number |
663100 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a automate tool refill broke during use.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Adding udi # : udi # (b)(4).Returned product was 1 automate iii handle and 1 flexshaft date code 0622 with coil deformation/weld failure causing the product to not function properly.Capa-2021-375 opened to address weld failures for product manufactured by sarasota since december 2020 (date code 1220) through implementation date of (b)(6) 2022 (0822).Complaint is considered substantiated.(nwv).
|
| |
|