| Model Number |
8884721252 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported the tube was inserted into the patient and the stylet wire was discarded.The wire punctured the doctor's thumb during removal.Additional information received stated the green cap covering the end of the wire fell off and the wire punctured the doctor's thumb.The doctor left the case, washed his hands, and applied a band aid.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h4 device manufacture date was added.The device history record (dhr) was reviewed showing no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.Sample evaluation could not be performed because no sample nor photo was available for evaluation.A walk through was performed of the manufacturing process.Process and controls were found properly followed.No action plan is required at this point.The manufacturing site will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.
|
| |
|