| Model Number |
461055E |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported after two days of use, the port disconnected from the rubber y it attaches to.The y was leaking where it connects to the tube.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review could not be performed because the serial number was not available.The provided photo indicates the y port cap detached from the body.The root cause of the reported issue could not be determined.There is no risk of y-port damage during manufacturing as testing and inspection can 100% detect y-port damage during the assembly process.No action will be taken at this time.This complaint will be used for tracking and trending purposes.
|
| |
|