| Model Number |
8884710859E |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported an 8 french dobhoff tube was placed in the patient, and it was verified by xray.They were unable to withdraw the guidewire.A second attempt was made to place the dobhoff, which was verified by xray, but still was unable to pull the guidewire.The night rn attempted to remove the guidewire but was unable to so they replaced it with a 12 french dobhoff and successfully were able to pull the guidewire.Order in for verification of placement to begin tube feedings.The patient tolerated the 3rd attempt well.Tube feedings to continue after verification of placement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed showing that manufacturing and inspection of product was performed as per applicable procedures and validated process.One sample was received for evaluation.The hydromer was activated with approximately 10ccs of water, and the stylet was removed without problem.A failure to activate the hydromer would make it difficult to remove the stylet from the feeding tube.The investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly.No action plan is required at this point.The manufacturing site will continue to monitor customer complaints and feedback notifications for adverse trends that require immediate attention.
|
| |
|
Search Alerts/Recalls
|