• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FATHOM -14; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FATHOM -14; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 81725
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device fracture occurred.The target lesion was located in the severely tortuous and moderately calcified vessel.Two 200 cmx10 cm fathom-14 guidewire were selected for use.During advancing, it was noted that device fracture occurred approximately 6 inches from the top of the wire.The first device was snared and since this device was still connected, it was easily removed.As per physician's opinion, the fracture was possibly caused by the tortuosity of the vessel and going through stent.The procedure was completed with a different device.There were no complications reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device and a torque device were returned for the analysis, and it is possible to see that the guidewire was kinked and detached at the distal section.No more damages were detected.Microscope inspection revealed that under the microscope it was observed that the guidewire was kinked and detached at the distal section.
 
Event or Problem Description
It was reported that device fracture occurred.The target lesion was located in the severely tortuous and moderately calcified vessel.Two 200 cmx10 cm fathom-14 guidewire were selected for use.During advancing, it was noted that device fracture occurred approximately 6 inches from the top of the wire.The first device was snared and since this device was still connected, it was easily removed.As per physician's opinion, the fracture was possibly caused by the tortuosity of the vessel and going through stent.The procedure was completed with a different device.There were no complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FATHOM -14
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17079630
Report Number2124215-2023-27092
Device Sequence Number2173688
Product Code DQX
UDI-Device Identifier08714729801023
UDI-Public08714729801023
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K111498
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2026
Device Model Number81725
Device Catalogue Number81725
Device Lot Number0031256293
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/16/2023
Supplement Date Received by Manufacturer06/22/2023
Initial Report FDA Received Date06/07/2023
Supplement Report FDA Received Date07/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-