Product # pc-001355222 date sent to the fda: 6/7/2023, component code: g07002.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and no answer was obtained.If any additional information is received, a supplemental medwatch report will be sent."could you please provide the lot number for d9782 product code? device return status.Please document the shipment tracking number: please provide the source or name of person providing answers to follow-up questions (not the person relaying/submitting answers to loc or chu)." attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent related reports: 2210968-2023-04152.
|
|
Product complaint # (b)(4).Date sent to the fda: 7/5/2023.H6 component code: g07002 no device problem found.Additional information: d4, d9, h3, h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.The returned sample determined that it was received one labeled winding former with one detached needle that pertain to the product code 1611g.Upon visual assessment of the sample, the suture was not returned for analysis.The needle hole and swage area were noted with marks that appear to be caused by a surgical instrument.The barrel hole was examined under 20x magnification, and no suture remnant could be observed.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.No further investigation will be conducted on this complaint due to external cause.Complaint information will be included in complaint trending that is reviewed by the applicable product quality safety surveillance data review board on a regular basis to determine if further action is necessary.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample determined that it was received one unopened sample that pertain to the product code 1611g.Upon initial inspection, of the samples, no external damages were observed on the packet.In order to evaluate the condition of the returned sample, the packet was opened.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-04152.
|