Catalog Number MP5314-C
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd maxplus pressure rated extension set with removable needleless connector was missing component.The following information was provided by the initial reporter: mtm2 found extension set missing one of the ends during production.
 
Additional Manufacturer Narrative
B.3.Date of event: unknown.The date received by manufacturer has been used for this field.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.D4.Medical device lot #: the customer reported, lot #: 22079077, but this was not found to be associated with the reported material number.
 
Event or Problem Description
It was reported that the bd maxplus pressure rated extension set with removable needleless connector was missing component.The following information was provided by the initial reporter: mtm2 found extension set missing one of the ends during production.
 
Additional Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 22-jun-2023.H.6.Investigation summary: two hundred six samples of the mp5314-c tubing and 155 maxplus connectors were received for quality investigation.Two samples of the 206 samples were seen to be missing the male luer connector or both the male and female luer connectors.The customer complaint of misassembly was confirmed by visual inspection.Further examination of the two samples shows that there is an absence of solvent at the end of each tubing.The additional tubing samples submitted did not have the maxplus connectors connected to the assemblies.The tubing samples have male and female connectors on securely connected each end.The 155 samples of the maxplus connectors were not assembled on the tubing and provided in a separate plastic bag from the tubing assemblies.Connecting the maxplus connectors to the extension set tubing and pushing water through the full assembly using a water filled syringe, did not indicate any flow issues, connection issues or air-in-line issues.After investigation, the main contributor of misassembly/missing of the female and male luer could be related to not following the assembly instructions and incorrect segregation of suspect material by the assembler.A quality alert has will be generated to reinforce the correct assembly of the model and the correct segregation of the suspect material to avoid this issue in future production.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXPLUS PRESSURE RATED EXTENSION SET WITH REMOVABLE NEEDLELESS CONNECTOR
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17081232
Report Number9616066-2023-01129
Device Sequence Number2137131
Product Code FPA
UDI-Device Identifier10885403236723
UDI-Public(01)10885403236723
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K083472
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMP5314-C
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/16/2023
Supplement Date Received by Manufacturer10/13/2023
Initial Report FDA Received Date06/07/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-