Model Number N160
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that technical services (ts) was contacted regarding crosstalk oversensing being flagged by a third party reviewer, despite it falling within the device's blanking period.Ts discussed the phenomenon with the physician, and they both agreed the crosstalk was being appropriately blanked by the device.The device and leads remain in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that technical services (ts) was contacted regarding crosstalk oversensing being flagged by a third party reviewer, despite it falling within the device's blanking period.Ts discussed the phenomenon with the physician, and they both agreed the crosstalk was being appropriately blanked by the device.The device and leads remain in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17081573
Report Number2124215-2023-29155
Device Sequence Number17703174
Product Code LWP
UDI-Device Identifier00802526480997
UDI-Public(01)00802526480997(17)161113(21)969850
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/13/2016
Device Model NumberN160
Device Catalogue NumberN160
Device Lot Number969850
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/23/2023
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date06/07/2023
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
-
-