| Model Number |
N160 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that technical services (ts) was contacted regarding crosstalk oversensing being flagged by a third party reviewer, despite it falling within the device's blanking period.Ts discussed the phenomenon with the physician, and they both agreed the crosstalk was being appropriately blanked by the device.The device and leads remain in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that technical services (ts) was contacted regarding crosstalk oversensing being flagged by a third party reviewer, despite it falling within the device's blanking period.Ts discussed the phenomenon with the physician, and they both agreed the crosstalk was being appropriately blanked by the device.The device and leads remain in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Search Alerts/Recalls
|