| Model Number |
2092-6040 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block a2: the patient''s dob is unknown.This information was not available from the facility.Block b7: patient information regarding medical history is unknown.This information was not available from the facility.Block g2: foreign- japan.Block h3: the angiosculpt device was discarded by the facility, thus no returned product investigation was performed.Block h6: the ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
The angiosculpt device was used to treat the distal sfa.During the first inflation at 6 atm for 5 seconds, the balloon ruptured.The procedure was completed with a non-angiosculpt device.No patient injury reported.This product problem is being submitted conservatively because the balloon ruptured below rbp.
|
| |
|
Search Alerts/Recalls
|