A1: patient identifier: requested, not provided a2: age & date of birth: requested, not provided a3: patient sex: requested, not provided a4: weight: requested, not provided a5: ethnicity: requested, not provided a6: race: requested, not provided d6a: implanted date: device was not implanted d6b: explanted date: device was not explanted e1: initial reporter name : requested, unknown e2: health professional: requested, unknown e3: occupation: requested, unknown the actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.History investigation of the involved product code/lot #.Since the involved lot was unknown, the investigation of manufacturing history records and shipping inspection records could not be performed.History investigation of the involved product code.There have been no similar complaints in the past two years from other facilities about the involved product code.Since the involved lot number was unknown, only di no.Is listed.Since the involved lot was unknown, it was not possible to investigate the manufacturing records and shipping inspection records.Since the actual sample was not returned and investigation could not be performed, it was not possible to clarify the cause of occurrence.The instructions for use (ifu) of this product includes the following warning.Warning - if any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the glidewire and/or the catheter and determine the cause by fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the guide wire's tip, damage to the catheter, or damage to the vessel.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.
|
|
The user facility reported that during a cryo atrial fibrillation procedure, an effusion occurred in the inferior vena cava (ivc) which required a vascular balloon.There were difficulties while trying to gain access into the right atrium at the beginning of the procedure.Eventually all accesses were made in the proper locations.After freezing the left veins and moving to the right side, the patient's pressure was dropping, and heart rate was increasing.Nothing observed with the intracardiac echocardiography (ice) nor the transesophageal echocardiography (tee).After checking for effusion, the physician injected contrast low in the ivc looking for a perforation and found a small one below the heart just below the diaphragm on the lateral side of the ivc.Likely the perforation was due to one of the catheters during access.A vascular balloon was inflated in ivc with stent placed to seal the perforation.The case was abandoned before any rf was done.The patient was stabilized, and no further complications occurred.There were no alleged performance issues with any abbott devices.There were no catheters in ivc.It is alleged that the guidewire punctured the ivc at beginning of procedure.Procedure performed was an atrial fibrillation.The event occurred intra-operative.There was no reported blood loss.The patient was not injured during the event and medical or surgical intervention was not required.There was not a direct allegation that the reported device caused or contributed to patient injury and/or need for medical intervention.
|