|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd drive pcb.In addition, our technician confirmed that the root brace rubber (insertion flexible tube) cut, the ccd drive pcb malfunction, the distal body worn out, the light guide cable lump/wave, the suction channel kink, and the up pulley wire snapped/cut.However, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|