The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd drive pcb.In addition, our technician confirmed that the bending rubber leak, the u/d and the r/l knobs loose, the light guide cable lump/wave, the angulation down angulation decrease, the angulation left angulation decrease, the angulation right angulation decrease, and the angulation up angulation decrease.However, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|