| Model Number |
7711700 |
| Medical Device Problem Codes |
Fracture (1260); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
Air Embolism (1697)
|
| Date of Event |
05/12/2023
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that five months and eight days post the chronic catheter placement, an alleged crack was observed on the catheter valve connector causing air embolism.It was further reported the patient expired.Reportedly, the patient was being treated for severe pulmonary arterial hypertension and epoprostenol was being administered.Patient's cause of death is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: date of death for this patient was not provided, date of death updated as (b)(6) 1900 for the mdr submission requirement.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 05/2026).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that sometime post a chronic catheter placement, an alleged crack was observed on the catheter valve connector causing air embolism.It was further reported the patient expired.Reportedly, the patient was being treated for severe pulmonary arterial hypertension and epoprostenol was being administered.Patient's cause of death is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failures of catheter connector crack and attendant air embolism as no objective evidence was provided for review.Note that the connector was described as a "groshong valve connector." to clarify, the connector itself does not have a valve, but the tip of the catheter does.Also, the relationship between the reported catheter connector crack/air embolism and the patient death has not been established.The definitive root cause could not be determined based upon the available information.Labeling review: as the reported event did not allege a labeling or use related issue, therefore a labeling review is not required.H10: d4 (expiration date: 05/2026), g3, h6 (method) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|