| Model Number |
M00562321 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Failure to Cut (2587)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a captivator medium oval stiff snare was used during a polypectomy procedure performed on an (b)(6) 2023.During the gastric polyp resection procedure in the ascending colon, the medical team encountered difficulty in using the captivator single-use polypectomy snares due to its small ring size.After the device was opened, it was observed to be in a strip shape.Following the cold incision, the wound began to bleed profusely and could not be stopped with irrigation.The bleeding was eventually stopped by the application of a clip to the wound and the procedure was completed with a similar captivator medium oval stiff snare.It was not reported if there were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block e1 (initial reporter address 1) (b)(6); block h6: imdrf device code a050702 captures the reportable event of loop cutting issues.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a captivator medium oval stiff snare was used during a polypectomy procedure performed on (b)(6) 2023.During the gastric polyp resection procedure in the ascending colon, the medical team encountered difficulty in using the captivator single-use polypectomy snares due to its small ring size.After the device was opened, it was observed to be in a strip shape.Following the cold incision, the wound began to bleed profusely and could not be stopped with irrigation.The bleeding was eventually stopped by the application of a clip to the wound and the procedure was completed with a similar captivator medium oval stiff snare.It was not reported if there were no patient complications reported as a result of this event.Per additional information received on july 12, 2023, it was reported that cold incision was performed to resect a 1cm polyp, even though the snare loop did not fully extend, and it was the original device that was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Block e1 (initial reporter address 1): (b)(6).
|
| |
|