| Model Number |
IPN039296 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while in use, the belly bag is "leaking from where the inlet tube meets the bag".Additional information received states that there was no patient harm or injury, or medical intervention required.The issue was resolved by changing the bag.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that while in use, the belly bag is "leaking from where the inlet tube meets the bag".Additional information received states that there was no patient harm or injury, or medical intervention required.The issue was resolved by changing the bag.The patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|