• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NEFF PERCUTANEOUS ACCESS SET; KGZ ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NEFF PERCUTANEOUS ACCESS SET; KGZ ACCESSORIES, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Remove (1528); Unraveled Material (1664); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the wire guide included in the neff percutaneous access set became stuck in the puncture needle and began to unravel during a choledochal procedure on an 87-year-old male patient.After the wire guide became stuck in the needle, it became difficult to advance or withdraw.Following an overall withdrawal from the patient, the product was checked.The user noted that the wire was, 'slightly cut," however the device was continued to be used and the bile duct was successfully punctured again.The patient did not require any additional procedures or experience any adverse effects due to this occurrence.Additional information regarding event details has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
E1 - customer (person): facility - (b)(6) hospital.G2 - report source - other: competent authority nmpa.G4 - pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Additional information received from the customer clarified that the wire guide used in the event did not unravel.There is no evidence to suggest that the observed device failure would be likely to cause or contribute to a death or a serious injury if the failure were to recur.Furthermore, there is no evidence that the device caused or contributed to a serious injury, as no life-threatening or permanently impairing injury took place, nor was intervention taken as a result of the device failure which would be required to prevent permanent impairment or damage to the patient.As such, the event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury, or reportable product malfunction.No further action is required and no additional follow ups will be sent for this event.
 
Event or Problem Description
In additional information received from the customer 14jun2023, it was reported that the wire guide did not unravel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEFF PERCUTANEOUS ACCESS SET
Common Device Name
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key17088483
Report Number1820334-2023-00705
Device Sequence Number14995557
Product Code KGZ
UDI-Device Identifier00827002105447
UDI-Public(01)00827002105447(17)250429(10)14699904
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/29/2025
Device Model NumberN/A
Device Catalogue NumberNPAS-100-RH-NT
Device Lot Number14699904
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/05/2023
Supplement Date Received by Manufacturer06/14/2023
Initial Report FDA Received Date06/08/2023
Supplement Report FDA Received Date06/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age87 YR
Patient SexMale
-
-