| Model Number |
MEDISORB, MULTI ABSORBER ORIGINAL |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not yet returned.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the bellows do not stay inflated after changing the medisorb, multi absorber original during procedure.The leak rate is greater than 750ml.The soda lime container was replaced which resolved the issue.The patient was not harmed.
|
| |
|