Model Number LOGIC TIBIA PS MOD INSRT SZ 2.5 11MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377); Implant Pain (4561)
Date of Event 12/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: (b)(6), 02-010-01-0225 - logic femoral ps cem left sz 2.5; (b)(6), 02-012-45-2515 - lgc tibial fit tray cem sz 2.5f / 1.5t; (b)(6), 200-02-32 - three peg patella 32mm; (b)(6), 204-34-02 - fluted stem extension 25l x 14 mm; (b)(6), 204-70-00 - tibial stem ext.Screw.Pending investigation.
 
Event or Problem Description
As reported via legal department 64 y/o female patient had a left knee replacement on (b)(6) 2013.Approximately 7 years and 4 months after the initial procedure the patient had a left knee revision on (b)(6) 2020.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.Revision op report on (b)(6) 2020.Post-op diagnosis: failed left total knee arthroplasty secondary to osteolysis.Aseptic loosening femoral component left total knee.Large cortical osteolytic defect non-contained anterior medial tibia.Four large area of osteolysis in the lateral distal femur, contained defect.Findings: loose femoral component with de bonding from the implant between the implant and the cement mantle non-contained anterior medial tibial defect secondary to osteolysis severe polyethylene wear from the medial side of the polyethylene tray, large lytic defect in the anterior distal femur laterally this was a contained defect severe thinning of the lateral femoral condyle.The patient tolerated procedure well and transferred to recovery in stable condition.
 
Additional Manufacturer Narrative
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.H6: corrected health effect, component, and investigation clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIBIA PS MOD INSRT SZ 2.5 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, CEMENTED, POLYMER/METAL/POLYMER
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
gainesville 32653
3523771140
MDR Report Key17088767
Report Number1038671-2023-01292
Device Sequence Number6070524
Product Code JWH
UDI-Device Identifier10885862001771
UDI-Public10885862001771
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K093360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/29/2015
Device Model NumberLOGIC TIBIA PS MOD INSRT SZ 2.5 11MM
Device Catalogue Number02-012-35-2511
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2023
Supplement Date Received by Manufacturer02/10/2025
Initial Report FDA Received Date06/08/2023
Supplement Report FDA Received Date02/10/2025
Date Device Manufactured02/01/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient RaceWhite
-
-