| Model Number |
WA47786A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/04/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device referenced in this report was not returned to olympus for evaluation.The root cause cannot be determined at this time.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
A user facility reported to olympus that while using the versapoint bipolar electrode resectoscopic - 2.5 mm angled loop, resecting, the distal end of the loop broke in the patient 30 minutes into the procedure.The loop was not found.The procedure continued with a second loop, which also broke.The procedure was then abandoned/interrupted.The patient was given prophylactic antibiotics due to the potential infection risk.It was stated that the physician did not feel there was a risk of uterine perforation or any other serious injury.Additional information was requested, but not provided.This event requires two reports for the two broken loops.Patient identifier's (b)(6) and (b)(6).This report is for (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.The loop at the distal end of the item was broken.The product was forwarded to the supplier, whom also confirmed the reported issue.The active tip of the device was fractured from the distal tip.No electrical or functional testing was conducted.The condition of the fracture face of the device suggests a brittle failure.The results of the previously reported investigation do not change.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause for this issue is very likely weakness in current material strength and ductility.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was additionally reported that the procedure was a hysteroscopy and that the patient had a good post surgical recovery.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information obtained from the customer regarding the reported event.New information added to the following fields: b1, b5 and d4 (expiration date).
|
| |
|