Catalog Number 115397
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Fall (1848); Failure of Implant (1924); Foreign Body In Patient (2687)
Date of Event 09/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01317, 0001825034-2023-01318, 0001825034-2023-01319.Concomitant medical products: item# 180552; lot# 599180, item# 180552; lot# 966580, item# 180550; lot# 050440, item# 010000589; lot# 410120, item# 115310; lot# 795470, item# 00434903606; lot# 63953099, item# 00434901213; lot# 64351214.G2: foreign - event occurred in the uk.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a shoulder arthroplasty approximately three (3) years and nine (9) months ago.Subsequently, the patient fell on an unknown date and fractured screws within the baseplate.Patient was later revised approximately nine (9) months ago and converted to a hemi-arthroplasty.During the revision, it was noted three of the fractured screw shafts could not be removed from the bone and were retained by the patient.No adverse consequences have been reported as a result of the event.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.It was reported that a patient fell and sustained a fracture of the implanted screws.Prior to the fall, there were no allegations against or concerns with the implants.As upper extremities are not load bearing for balance or ambulation, the shoulder implants can be ruled out as causing or contributing to the external trauma of the fall which resulted in implant fracture and subsequent screw tip retention within the bone.During the revision, the surgeon chose to leave the lodged screw fragment in place to reduce further trauma or bone loss within the shoulder.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS CNTRL 6.5X35MM ST/RST
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17090546
Report Number0001825034-2023-01320
Device Sequence Number12279588
Product Code PHX
UDI-Device Identifier00880304677098
UDI-Public(01)00880304677098(17)290418(10)645890
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number115397
Device Lot Number645890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer09/27/2023
Initial Report FDA Received Date06/08/2023
Supplement Report FDA Received Date10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-