• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC HERCULES 3 STAGE BALLOON ESOPHAGEAL; KNQ, DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC HERCULES 3 STAGE BALLOON ESOPHAGEAL; KNQ, DILATOR, ESOPHAGEAL Back to Search Results
Model Number G48732
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with no sign of damage to the balloon or catheter.During our function test a water filled ds-60cc from our stock was attached to the inflation port.When injected three streams of water could be seen coming from pin holes in the middle of the balloon material.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation identified a pinhole leak in the balloon material.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The report states it is unknown if lubrication and negative pressure were applied to the balloon prior to advancement through the endoscope.A possible contributing factor to balloon material damage is failure to lubricate the balloon with a lubricating agent.The instructions for use direct the user: "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." this activity will aid in endoscopic advancement and balloon preservation.Another possible contributing factor to a leak in the balloon material is failure to apply negative pressure to the balloon dilator prior to advancement through the endoscope.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use direct the user "to facilitate passage through the endoscope, apply negative pressure to the catheter." in addition, the user is further instructed to "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." a possible contributing factor to a leakage in the balloon material is if the balloon comes in contact with a sharp object or burr in the endoscope accessory channel.Prior to distribution, all hercules 3 stage balloons esophageal are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a transendoscopic esophageal dilation, the physician used a cook hercules 3 stage balloon esophageal.It was reported [that] the user attempted to inflate the balloon at target site of esophagus, but it didn't inflate.He observed that the saline with contrast medium was leaked from the balloon [subject of report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HERCULES 3 STAGE BALLOON ESOPHAGEAL
Common Device Name
KNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key17091456
Report Number1037905-2023-00285
Device Sequence Number2833486
Product Code KNQ
UDI-Device Identifier00827002487321
UDI-Public(01)00827002487321(17)251209(10)W4663794
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2025
Device Model NumberG48732
Device Catalogue NumberHBD-15-16.5-18
Device Lot NumberW4663794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2023
Initial Report FDA Received Date06/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE - UNKNOWN MAKE AND MODEL
Patient SexUnknown
-
-