Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Retinal Detachment (2047); Loss of Vision (2139)
Type of Reportable Event Serious Injury
Event or Problem Description
An online article reported a case study with akreos opacification associated with 5000-centistoke silicone oil.The article reports a patient with a clinical diagnosis of subluxated intraocular lens (iol) with very poor zonular support was made.The patient underwent pars plana vitrectomy, single-piece iol removal, and scleral fixation of an akreos iol with gore tex sutures.Large holes developed during vitrectomy at 11, 1, and 7 o¿clock.The areas were secured with laser.Nine days post surgery, the patient experienced vision loss, and a peripheral retinal detachment was noted.An inferior peripheral iridotomy was created, and air was left in the anterior chamber.The next day, vision was 20/150 and the retina was attached.In addition, a small amount of fibrin could be seen on the anterior surface of the lens and a small air bubble remained in the anterior chamber.After 3 months, visual acuity decreased to 20/800 despite the retina being flat and the oct not showing any macular structural abnormalities.Possible opacification of the iol was noted.During the planned silicone oil removal, attempts to scrub the surfaces of the iol were not successful and opacification of the lens itself was confirmed.The iol was explanted and replaced with a different model anterior chamber iol.
 
Additional Manufacturer Narrative
Article website: https://www.Canadianjournalofophthalmology.Ca/article/s0008-4182(17)31251-6/fulltext.The lot number number is not available.Therefore a review of the device history record could not be performed.Investigation of this event is ongoing and a supplemental report will be submitted upon completion of investigation.
 
Additional Manufacturer Narrative
Updated d4, g3, h2, h6, h10 the device was not returned for evaluation and requests for additional information was not fulfilled.The trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADATO SIL-OL 5000 SILICONE OIL SYRINGE
Common Device Name
FLUID, INTRAOCULAR
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
PHARMPUR GMBH
messerschmittring 33
konigsbrunn 86343
GM   86343
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key17091879
Report Number0001313525-2023-70069
Device Sequence Number2906119
Product Code LWL
UDI-Device Identifier00757770516315
UDI-Public(01)00757770516315
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P910071
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/08/2023
Supplement Report FDA Received Date08/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AO60 IOL.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexMale
-
-